Going beyond standard quantum limit with an optical spring

Author: Artem Basalaev

At the 10 m Prototype, several technologies will be used to reach incredible displacement sensitivity of the main interferometer. Our 100 g test masses will be coated with AlGaAs coatings and suspended by the thin silica wires from seismically-isolated optical tables. This setup will allow us to reach the Standard Quantum Limit (SQL) in sensitivity: the lower limit on the combination of the quantum radiation pressure noise (QRPN) and shot noise (SN) at each frequency, as dictated by the Heisenberg uncertainty principle. Achieving this level of sensitivity is necessary to create an environment that resembles current and planned future gravitational wave detectors to subsequently test new technology for them.

One focus is on developing technologies that allow us to surpass the SQL, that can be used in future detectors. One such technique is squeezing (injection of a squeezed vacuum state). This introduces a cross-correlation between QRPN and SN which reduces total achievable quantum noise. This technique has been successfully applied at current detectors, especially at LIGO, where the frequency-dependent squeezing has been perfected [Ganapathy et al 2023]. Frequency-dependent squeezing requires a high-finesse, low-loss, long-length filter cavity. At LIGO, the filter cavity is 300 meters long. At future Einstein Telescope (ET) detector, the filter cavity will have to be an incredible 5 km long to achieve 15 dB of squeezing in the low-frequency interferometer [Korobko 2025]. Needless to say, building such a long filter cavity with the required low loss and high finesse is extremely technically challenging.

Fortunately, there is another complementary technique that will help us surpass the SQL threshold for ET: the so-called optical spring [Corbitt 2007]. ET’s signal recycling cavity will be detuned from its resonant condition, introducing two distinct minima in the sensitivity spectrum: one corresponding to the optical resonance of the detuned signal recycling cavity, and another corresponding to the opto-mechanical resonance of the optical spring. Carefully chosen detuning ensures the highest sensitivity at low frequencies up to around 20 Hz. At higher frequencies the high-frequency interferometer of ET will take over.

The optical spring amplifies the amplitude of the classical output signal at the optical spring frequency, while leaving the output quantum fluctuation unaffected. Therefore, it can and will be combined with frequency-dependent squeezing to maximize benefit. However the two-minima structure of the sensitivity, created by the optical spring, requires not one but two 5-km-long filter cavities for squeezing! Nevertheless, the benefits of the optical spring are expected to outweigh the challenges. Therefore the two-filter-cavity, optical spring, combined with frequency-dependent squeezing, is currently the envisioned configuration for the low-frequency interferometer of ET [Korobko 2025].

Unlike frequency-dependent squeezing, the optical spring has not yet been tested for shaping sensitivity beyond the SQL at the current gravitational wave detectors. Therefore it is a good candidate to test first at the 10 m Prototype to ensure the success of the future gravitational wave detectors, particularly ET.

Optical spring: how it works

Left: Fabry-Perot Michelson interferometer (such as the 10 m Prototype). Right: dual-recycled Fabry-Perot Michelson interferometer (e.g. LIGO, ET).

Creating an optical spring, for a signal-recycled Fabry-Perot Michelson interferometer, can be described as follows:

  • The signal recycling cavity is detuned from the resonance condition by moving the signal recycling mirror → affecting position of the input test masses and therefore the the field amplitude in the arm cavities.
  • The position of the arm cavity mirrors is also influenced by the radiation pressure force ← which depends on the field amplitude in the cavity.
  • This creates an opto-mechanical feedback mechanism that acts similarly as if the mirrors were connected by a mechanical spring (or an anti-spring).
  • At the optical spring frequency the sensitivity of the interferometer is resonantly enhanced. This effect is independent of the SQL, meaning that the SQL can be surpassed!

Optical spring at the 10 m Prototype

In a signal-recycled Fabry-Perot Michelson interferometer, an optical spring is created by moving the signal recycling cavity mirror. However, signal recycling is not planned for the 10 m Prototype. Nevertheless, all necessary conditions to introduce an optical spring will be met. Sub-SQL interferometer will reach the SQL in the range of approximately 65 Hz to 200 Hz, making it a good test case for real gravitational wave detectors. Moreover, unlike squeezing techniques, which shape the noise, the optical spring effect, which shapes the signal response, actually does not rely on being limited by the SQL.

Rather than using a signal recycling cavity, the optical spring will be introduced directly in the arm cavities by moving the end test masses, modifying the common arm (CARM) degree of freedom, defined in the schematic above. We will use our electrostatic drives (ESDs) to add detuning to the arm cavities by moving the end test masses. Typically, ESDs would keep the end test masses positioned such that the arm cavities are in a resonant condition. Here, however, we will use them to introduce a static offset of a few nanometers. This offset must be the same in both arms; otherwise, it will also change the DARM degree of freedom, which is undesirable because we will monitor the change in sensitivity there.

A study with Finesse 3

A study using realistic parameters for optics, geometry and the test mass suspensions model of the 10 m Prototype has been conducted with Finesse 3. The study was performed in two variations:

  • One used 5 W of main laser power (the default configuration foreseen at the 10 m Prototype). The end test masses were moved by up to 4 nm, corresponding to detunings between -0.10 and -3.00 line widths of the arm cavity, with a step of 0.25 line widths.
  • The second used 50 W of main laser power (which would be possible if power recycling is introduced at the 10 m Prototype). In this case, the end test masses were moved by up to 9 nm, corresponding to detunings between -0.25 to -6.50 line widths of the arm cavity, with a step of 0.10 line widths.

The points of maximum sensitivity from this sweep measurement were connected, to create an optical spring sensitivity limit. The result is shown in the figure below.

The effect of the optical spring in the CARM degree of freedom for the 10 m Prototype (sweep noise and two examples of specific detunings). Left: default 5 W of laser power; right: 50 W of laser power, which is possible if power recycling is introduced. The shaded gray area represents 65-200 Hz window, in which the 10 m Prototype is expected to be limited by the SQL. Outside of this window, classical noise sources will dominate.

The study’s conclusion is that, with the anticipated 5 W of laser power, the optical spring’s improvement should be detectable at various detunings in CARM. For 50 W of laser power, the improvement is more significant but requires larger detunings. The complete simulation, together with the exact parameters of the 10 m Prototype and its suspensions is available in the attached Jupyter notebook.

From the Finesse 3 study to implementation

An attractive property of the optical spring experiment at the 10 m Prototype, is that it does not require any additional changes to the planned Fabry-Perot Michelson interferometer. The detuning in CARM will be achieved using the ESDs that will be installed for controlling the test masses. Monitoring will be performed in the DARM degree of freedom. There’s no conflict with any other planned experiments, that can be set up in parallel. The other experiments will only be affected during the short measurement time when the end test masses are actively moved. The optical spring effect is also robust against unwanted disturbances. Unlike techniques manipulating quantum noise, it can be observed even when classical noise sources dominate. Lastly, even a fully functional Fabry-Perot Michelson interferometer is not required. Initial studies can be conducted with a single cavity.

An important factor for successfully quantifying the impact of the optical spring is precise calibration of the detuning. It is important to move the end test masses to the same detunings, thereby changing the CARM degree of freedom. Any mismatch will also introduce a DARM offset. From the variations of the parameters in the Finesse 3 study linked above, it was found that the optical spring can tolerate a small DARM offset. However, an unknown offset will complicate the interpretation of the results.

To calibrate the ESDs, a photon calibrator (PCAL) will be used. A PCAL is a laser that applies precisely known radiation pressure to the mirror. The produced force can be compared to the ESD and therefore used to calibrate it. To calibrate the PCAL itself, an integrating sphere can be used. An integrating sphere precisely measures laser power by capturing nearly all laser light, which is then detected by a photodiode. The photodiode measurement is precisely calibrated by the manufacturer, with calibration traceable to a NIST or PTB standard.

Calibration should also be monitored to compensate for possible drifts during the measurement by the means of injecting additional calibration lines. Both ESDs will be excited at around 500 Hz and monitored in DARM. Any difference in the observed amplitudes of the lines in DARM will therefore indicate miscalibration and will be corrected by adjusting the ESD gains. At lower frequencies, below 10 Hz, another set of two lines can be injected with the ESDs. At these frequencies, the actuation is mostly done by the coils on penultimate mass (PUM, the second suspension stage), which fully compensates for the induced low-frequency motion of the test masses. Therefore, the difference is that, for low frequencies, the calibration lines should be monitored in the PUM actuation signals instead of using DARM directly. Having these four calibration lines present ensures stability of the calibration.

Surfing the (gravitational) waves: dynamical signal tracking

As the study above showed, changing the cavity’s detuning shifts the optical spring frequency. Gravitational wave signals from inspiraling compact binaries follow a characteristic “chirp” pattern. As they progress toward merger, they rapidly rise in amplitude and frequency. Can we change the tuning on the fly, to maintain detector’s maximum sensitivity as the signal progresses? As if surfing gravitational waves? The answer is yes! However several challenges must be taken into account for such dynamical signal tracking.

First, feasible optical spring frequencies in a gravitational wave detector like ET tend to be at the lower end of the detection range, below a few hundred Hz. It may be possible to overcome this limitation by using more complex setups, such as utilizing multiple optical springs [Rehbein et al 2008]. However, it is more feasible to first focus on the advantages that an optical spring can offer in terms of low-frequency sensitivity.

Second, while the (static) optical spring is in the baseline of the ET low-frequency interferometer, there are additional challenges arise from dynamic changes in tuning. These challenges relate to controlling the interferometer and interpreting the results.

All current and planned gravitational wave detectors rely on linear control theory to operate. However, a Fabry-Perot Michelson interferometer is fundamentally a nonlinear system, where changes in power depend strongly, and nonlinearly, on the mirrors’ detuning. The trick that allows this to work is that, in the operational state, changes to the mirror positions are so small that changes in power considered linear. A linear controller is fully capable of maintaining desired positions. However, dynamical detuning to follow a progressing signal may no longer allow this. In fact, a state in which nonlinear dynamics are dominant is observed in current detectors all the time—during alignment and before the detector locks. Currently, this issue is overcome by a process that brings the detector to lock. This process often relies on random motion to bring the mirrors into the desired position by chance and then “catch” them there with a controller. This process is highly specific to each detector and often requires manual intervention. Fortunately, there has been significant progress lately in automated MIMO and nonlinear controls based on neural networks [Qin et al 2024; Ma, Vajente 2024]. A neural network can learn the nonlinear behavior of a system. These networks can be used as both sensors and controllers and can be combined with a Kalman filter to extract the maximum amount of information from limited and noisy inputs.

A second challenge arises from the behavior of noise during dynamical tuning. The noise floor is typically assumed to be stationary during the time that each signal spends in the detector. When this is not the case, we call it “glitches”, that have to be filtered out [Merritt et al 2021]. If this is not possible, the data may have to be discarded. Depending on the exact rate of change, with dynamic tuning, the noise may behave in a significantly non-stationary way. This may necessitate a reevaluation of the approach to data analysis and statistical inference. A changing noise model may need to be incorporated directly into the likelihood function used to infer the source parameters and their confidence intervals. Coincidentally, a similar situation may arise at ET even without dynamical tuning. Since ET will likely observe several signals simultaneously [Baka et al 2025], the other signals will serve as a dynamically changing background for any given signal. Therefore, developing robust inference methods for the dynamic case can generally be useful for future detectors.

It remains to be seen how much the above challenges will limit dynamical tuning for signal tracking. The 10 m Prototype experiment is ideally suited for this study. After setting up the optical spring with static tuning, dynamical tuning can be gradually introduced. A mock signal can be injected with the PCAL, and then recovered in DARM and reconstructed. The rate of tuning can then be increased, and the detector control and noise behavior can be closely monitored. This will allow us to find the threshold at which nonlinearities and nonstationary noise play a significant role and develop algorithms to address them.

If dynamical signal tracking proves reliable, new and exciting possibilities emerge. For example, highly eccentric black hole inspirals will manifest at low frequencies and feature a prominent second harmonic [Romero-Shaw et al 2019]. The optical spring can amplify this second harmonic to further boost confidence in the estimation of parameters of such a source. This would provide strong evidence that black holes coexist in dense environments such that they can encounter each other and merge within a short time, astronomically speaking. Furthermore, if the LISA space-based detector comes online, it will likely observe intermediate-mass black holes. Some of these black holes may transition into the ET observation band within minutes [Jani et al 2020]. This would allow us to anticipate their arrival and start dynamical signal tracking at precisely the right moment.

To fully utilize the potential of future gravitational wave detectors, we may need to rethink how we operate them. To do so, we must test new promising techniques, such as the optical spring and signal tracking with it, using available prototypes and simulations.